What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Reaffirmed January 2022

Revised January 2016, April 2009 

Originally approved October 2001, replacing "Expired Carbon Dioxide Monitoring" (September 1994)

Confirmation of proper endotracheal tube placement should be completed in all patients at the time of initial intubation both in the hospital and out-of-hospital settings. Physical examination methods such as auscultation of chest and epigastrium, visualization of thoracic movement, and fogging in the tube are not sufficiently reliable to confirm endotracheal tube placement. Similarly, pulse oximetry and chest radiography are not reliable as sole techniques to determine endotracheal tube location.

During intubation, direct visualization of the endotracheal tube passing through the vocal cords into the trachea, especially with the use of a videolaryngoscope, constitutes firm evidence of correct tube placement, but additional techniques should be used as objective findings to confirm proper endotracheal tube position.

Use an end-tidal carbon dioxide detector (i.e., continuous waveform capnography, colorimetric and non-waveform capnography) to evaluate and confirm endotracheal tube position in patients who have adequate tissue perfusion.

Esophageal detector devices are not as reliable as the various forms of capnography for the verification of endotracheal tube placement.

For patients in cardiac arrest and for those with markedly decreased perfusion, both continuous and non-waveform capnography may be less accurate. In these situations, if capnography is inconclusive, other methods of confirmation such as an esophageal detector device, ultrasound, or bronchoscopy should be used.

Ultrasound imaging may be used to reliably confirm endotracheal tube placement. However, this should be performed by someone who is experienced in this technique.

In recent years point of care ultrasound (POCUS) has been shown to be an exciting tool in the neonatal field, not only helpful in the diagnosis and follow up of a large number of clinical conditions, such as respiratory distress syndrome, transient tachypnea of the newborn, pneumonia, atelectasis, pneumothorax (1, 2) bronchopulmonary dysplasia (3) but also useful in devices' positioning and monitoring (4), thus constituting a promising radiation-free aid in clinical practice (5). Some reports explored POCUS application in endotracheal tube (ETT) position's assessment although strong evidence is still lacking (4), as demonstrated by the presence only of single center small sized studies and the absence of randomized controlled trials and of a systematic review and metanalysis (6). Endotracheal intubation is a very common procedure both in neonatal intensive care units (NICUs) patients facing respiratory failure, and in emergency settings like neonatal resuscitation in the delivery room (7). Endotracheal intubation is challenging in neonates and requires a high level of precision, due to narrow and short airways, especially in preterm newborns. Confirmation of the correct ETT placement is mandatory, since its malpositioning is quite common. Indeed a selective bronchial intubation is reported in 7% of cases (8) and can lead to severe complications such as atelectasis, airleaks, with consequent inadequate ventilation and possibly leading to morbidities and even mortality (9, 10). On the other hand an ETT placed too high could lead to accidental extubation. Esophageal intubation has a reported incidence of 21.4% (8) but is usually promptly recognized. Moreover, in case of respiratory distress syndrome, incorrect ETT location will lead to ineffective or inadequate surfactant administration (11).

Chest X-ray (CXR), currently considered the gold standard, is the most frequently used method for ETT's position confirmation in newborns (12, 13). However, it presents some limitation, such as ionizing radiation exposure (14) especially when additional CXR is required (for example in case of unintended extubation or need for tube repositioning) (15). Furthermore, CXR is time consuming, particularly in emergency settings, such as in the delivery room or prior to surfactant administration (7) like during an InSurE procedure (Intubation—Surfactant administration—Extubation), which requires fast and temporary intubation. In this setting, given the impossibility of performing CXR, end tidal CO2 (EtCO2) monitoring was proposed to assess correct positioning of ETT (16).

So, up until now there is no evidence on which is the most effective method for ETT position assessment in neonates (14). The ideal technique should be fast, non-invasive, radiation free, reliable, easy to learn and perform. The simplest, but unfortunately imprecise, way may be the evaluation of clinical signs, such as prompt improvement of heart rate and oxygen saturation, presence of thoracic excursions, equal breath sounds bilaterally in the lungs, condensation in the ETT (17). Capnography is another method, based on exhaled CO2, but can lead to both false-negative results, for example in case of low cardiac output or too low inflation pressure in presence of severe respiratory failure, and false-positive results in the event of right mainstem intubation (18). EtCO2 monitoring is also recommended by the “European Resuscitation Council” to confirm endotracheal tube position during advanced life support in newborns, even if it does not permit identification of selective bronchial intubation (13). POCUS has already been demonstrated to be a reliable technique for endotracheal tube confirmation in adults (19, 20) and in the paediatric population (21) while evidence is lacking regarding neonatal patients (13). Children's unique thoracic anatomy provides many acoustic windows into the chest allowing a suitable thoracic POCUS examination. Moreover the incomplete thoracic ossification of the newborn warrants a better analysis of portions of the anterior thorax and mediastinum (22). There has recently been a growing interest in performing POCUS in the NICU's setting, thanks to its advantageous technical features such as non-invasiveness, absence of radiation exposure, bedside feasibility, fast execution and repeatability. Data available in literature provided some different methods for visualizing the ETT tip and determining its position in the newborn population. Dennington et al. suggested placing the probe on the suprasternal notch to provide a midsagittal plane and, in order to confirm the tip is being seen, the ETT could be moved in and out (23), while in other studies a direct visualization of the ETT in the trachea is preferred (24). Other authors searched for the “comet tail” artifact (25) or the “double line” as confirmation of the tube in the trachea, while the superior right portion of right pulmonary artery or the aortic arch are the anatomic markers used for suggest the corrected ETT position (26, 27).

Objective

Primary objective: to evaluate POCUS capability in determining ETT placement adequacy in preterm and term newborns. Firstly, evaluating POCUS capability of identifying the presence of ETT in the trachea; then, its ability in detecting an appropriately positioned ETT as assessed by the gold standard technique. Secondary objective: to compare time required for POCUS confirmation vs. time needed for other standard of care methods (CXR and others).

Methods

Literature Searches

The study protocol was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) Statement.

A comprehensive search was conducted by three authors (SC, FS, MA) in Medline, PubMed, Google Scholar and the Cochrane Central Register of Controlled Trials (CENTRAL), from database creation to September 2021.

Manual searching of previous reviews and cross-references and contacting expert informants were undertaken to identify relevant papers to be analyzed. We also searched clinical trials registries for current and recently completed trials.

The search strategy included the following key words or Medical Subject Headings (MeSH) terms: “newborn”, “infant”, “neonate”, “endotracheal intubation”, “endotracheal tube”, “ultrasonography”, “ultrasound”; we adopted the Boolean operators “AND, OR” to connect our search words. We also used age filters by choosing the category “newborn” and considered papers published in the last 10 years (published after 2012).

Inclusion and Exclusion Criteria

Included studies met the following criteria: (1) NICU population needing intubation/intubated infants; (2) evaluation of POCUS efficacy and/or accuracy in detecting ETT position vs. a defined gold-standard method; (3) randomized and non-randomized controlled trials, prospective, retrospective papers and cross-sectional studies published in peer reviewed journals after 2012 were considered.

We excluded those studies of which, despite our best effort, we were unable to obtain the full text, mixed neonatal and pediatric population, case reports and case series.

Data collection and analysis was performed using the recommendations of the Cochrane Neonatal Review Group (28). Two authors (SF, AM) screened independently each article, reviewing the titles and abstracts of the studies and excluded studies that did not meet the eligibility criteria. The three review authors (SF, AM, CS) subsequently examined the retrieved full text of eligible studies.

Three review authors (SF, AM, CS) independently assessed the methodological quality of eligible studies and extracted data using a data extraction form blank data sheets are available in the Supplementary Material. In the data extraction form we reported for each eligible study: first author, year of publication, country, study design, type of enrollment, sample size and features, blinding, primary and secondary objectives, gold-standard method reported, POCUS operators and equipment, principal results. In case of incomplete data from a published paper the authors (SF, AM, CS) contacted the corresponding author requesting further information. Discrepancy in data extraction was resolved by discussion and building a consensus.

Quality Assessment and Risk of Bias

Quality assessment of each included study was evaluated independently by three review authors (SC, FS, MA) and cross-checked (FS for MA, MA for SC, SC for FS). Disagreements were resolved by discussion and consulting a fourth author (LB). Assessment of risk bias and applicability were performed using the QUADAS-2 tool (29), and the four key domains (patient selection, index test, reference standard, flow and timing) were investigated in each eligible study.

Data Synthesis and Statistical Analysis

To describe and summarize the data extracted by the different studies, we used dedicated tables. Furthermore, for each study (whereas such data were available) POCUS sensitivity and specificity to detect ETT position according to the gold standard technique (CXR) were represented in the forest plot; cumulative sensitivity and specificity values were also calculated through the contingency table. A secondary sub analyses was then performed to analyze these features in the subgroups using the two most common anatomic markers.

Statistical Software

RevMan software for Apple (5.4 version) was used for the statistical analysis including graphs of forest plot of sensibility and sensitivity and HRSOC curve. MedCalc.org was also used as an additional helpful tool.

Results

Study Selection

After a first screening based on title and abstract, and exclusion of duplicates, our electronic search for studies assessing the role of POCUS in the neonatal endotracheal intubation, provided a number of 18 studies. Full text of the studies was reviewed, and 4 other papers were excluded [1 due to mixed population and 3 were described as authors' experiences (30–33)]. Ultimately, 14 studies were included for data extraction and analysis. Flow chart of the search results, study selection log and included studies is presented in Figure 1, according to PRISMA guidelines (34).

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 1. PRISMA flow chart study selection.

Study Characteristics

Following the above specified criteria we selected 14 studies, including a total of 602 ETT evaluation with a range from 9 to 143 procedures per work. The studies were published from 2012 to 2021 and conducted worldwide: 4 in the USA (23, 26, 35, 36), 5 in Europe [Spain (11, 25), Italy (37), France (24), Russia (38)], 4 in Asia [India (39, 40), Japan (41), Iran (27)], 1 in South Africa (42). Table 1 provides the details of the studies included in the analysis, while Table 2 shows the methods' summary. Regarding the 14 studies included, a consecutive patients' enrollment was realized only in 4 cases (35–37, 40). In the other cases, the sampling method was not well declared, but we could understand that patients were enrolled according to the availability of the sonographer. Sampling method was not consecutive but random, based on the staff work shifts. In Table 1 we reported the studies' level of evidence using the Oxford grading system (43).

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Table 1. Details of the studies included in the analysis.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Table 2. Methods summary of the included studies.

Gestational age and birth weight of the newborns varied across the studies and were reported, where specified. 4 studies included extreme preterm infants <30 weeks (35, 37, 38, 41), 5 studies moderate preterm newborns 30-36 weeks (11, 23–25) and only one study (40) enrolled exclusively late preterm (>36 weeks). Mean birth weight of the patients included in the analysis, where clearly expressed, was >1,000 g, except in one case, where it was <1,000 g (41).

The inclusion criterion for the great majority of studies was need of intubation for any reason in patients admitted to NICU (23–27, 35–40, 42), with only two exceptions: in one case extremely low birth weight (ELBW) neonates were recruited when intubation was needed in the delivery room (41) and another one selected patients requiring surfactant therapy administered by InSurE technique (11).

Regarding exclusion criteria, when specified, patients presenting facial dysmorphism (26, 39) or upper airway or thoracic anomalies (23, 24, 26, 37, 39) were excluded for obvious reasons, as were patients with cardiac congenital defect (26, 35, 37, 39) since different position of aortic arch or right pulmonary artery would have interfered with ETT position estimation. Lack of parental consent was also considered (36).

The main objective of almost all the studies was to determine the efficacy of POCUS in assessing intubation and/or ETT position, in one case in relation to surfactant administration (11). Only one study had the specific goal of deriving normative data of the distance between ETT and anatomical structures across different weight and gestational age (39).

Regarding POCUS methods various markers have been proposed (Table 2). Images regarding probe position and ultrasonography projection are available in the Supplementary Material. When evaluating the mere tracheal intubation, 2 studies considered the comet tail artifact inside the trachea on cricoid transverse scan as marker of tracheal intubation (25, 41). On the other hand, when evaluating the precise position of ETT, the superior portion of the right pulmonary artery (RPA) was used as surrogate of carina anatomical position, while the apex of the aortic arch (AoA) was considered as a fixed point from which a distance of at least 0.5–1 cm was found to be appropriate. These POCUS scans were obtained from the suprasternal window with the probe placed in the midsagittal position. The distance between ETT tip and RPA was adopted by 4 studies (23, 26, 27, 37), while 7 studies measured AoA—ETT distance (24, 35, 36, 38–40, 42). Another study estimated the correct position assessing bilateral lung sliding and checking for the absence of acoustic shadow artifact in a longitudinal scan (corresponding to the air in the trachea if ETT was too high). For the other 2 studies the question about ETT exact position was not applicable (11, 41).

POCUS is performed by the figure of the neonatologist in half of the studies (11, 23, 24, 26, 27, 37, 41). Finally, POCUS equipment is quite heterogeneous in the different hospital centers of the studies included in our analysis, as showed in Table 2.

Quality Assessment

The results of the quality assessment, using QUADAS-2, of the risk of bias and applicability concerns of the selected studies are presented in Figures 2, 3.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 2. Risk of bias and applicability concerns graph according to QUADAS-2: review author's judgement about each domain presented as percentages across included studies.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 3. Risk of bias and applicability concerns summary according to QUADAS-2: review author's judgement about each domain for each included studies.

Considering the patient selection domain, all studies were considered to carry low risk of bias. Regarding those studies that enrolled patients according to the availability of the sonographer, the sampling method was random, therefore we do not believe that this could be a source of bias.

As for the index test domain, all studies have been judged as having low risk of bias except one (35), since POCUS was performed simultaneously with the reference standard (capnography) and the operator might have known the result of both the exams.

With respect to the reference standard domain, 3 studies (11, 40, 41) were considered at high risk of bias and 2 studies (26, 39) at unclear risk because the reference exam was inadequate and the blinding status was not explicitly reported, respectively.

With regard to the flow and timing domain, 2 studies (23, 36) were considered as high risk of concern because several hours elapsed between POCUS and CXR, moreover in one study not all the tests were reported in the final analysis; only one study was described as not clear risk.

Regarding applicability, 1 study (11) was considered to be at high risk of bias in relation to the reference standard domain, since the auscultation is described as the reference exam. With respect to the index test and patients selection domains, all studies were considered to have low concerns.

Role of POCUS in Detecting ETT in the Newborns

With respect to the data of the 14 selected studies, 602 ETT were investigated with POCUS. From the percentage numbers we derived data of interest (Table 3). In 583 cases (96.8%), POCUS identified the ETT.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Table 3. Number of intubation procedures described in the selected studies and POCUS efficacy in visualising ETT.

Accuracy of POCUS in ETT Placement in the Newborns

The diagnostic accuracy parameters (sensitivity, specificity, VPP, NPP) were derived by the results of the studies or, if not provided, from the raw numbers and vice versa using the Cochrane statistical package, RevMan V.5.4 software. Some studies were excluded from the accuracy analysis due to absence of specific data (26, 39) or different standard reference (e.g., capnography) (11, 41). Contingency table values, forest plot and HSROC curve regarding the pooled analysis are, respectively, reported in Table 4, Figures 4, 5.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Table 4. Pooled analysis contingency table.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 4. Pooled analysis forest plot: confirmation of ETT position by POCUS vs. CXR. ETT, endotracheal tube; POCUS, point of care ultrasound; CXR, chest X-ray; TP, true positive; FP, false positive; FN, false negative; TN, true negative; CI, confidence interval.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 5. Summary of the pooled HSROC curve.

The proportional metaanalysis revealed a pooled POCUS sensitivity of 93.44% (95% CI: 90.4–95.75%), intended as the POCUS ability in detecting an appropriately positioned ETT as assessed by the CXR. For those exams not capable of performing this task, the root cause was principally not sonography related, but due to malpositioned ETTs (ETT tip too high). In particular, Sethi et al. (40) were not able to detect ETT with POCUS in 5/48 patients, since the ETT tip was lying at or above the first thoracic vertebra; Najib et al. (27) could not see the ETT in 10/70 cases owing to it being higher than the thoracic inlet. On the other hand, few authors had difficulty in visualizing ETT by US due to poor quality of POCUS images: Singh et al. (39) in 10/143 patients; Descamps et al. (24) in 2/52. Moreover, one study (26) was excluded from the analysis, due to the absence of the data; authors were contacted but no response was obtained.

POCUS specificity was offered only by few studies because of the lack of true negative (TN) values (i.e., paucity of malpositioned ETT identified with the gold standard), the pooled POCUS specificity was 60.71% (95% CI: 40.58–78.5%). We obtained a PPV of 96.88% (95% CI: 95.15–98.01%) and NPV of 41,46% (95% CI: 30.35–53.58%).

We conducted a subevaluation regarding capability of POCUS in detecting a comparable ETT distance from carina compared with CXR. As regards the measurement of the distance between ETT and the superior aspect of the right pulmonary artery by POCUS, the data of 3 studies (23, 27, 37) can be extrapolated from the Bland Altman plot. If we consider +/– 0.75 cm as an acceptable deviation measure with respect to the depth of the ETT measured with the CXR, 93% of the measurements were in agreement. As arbitrary as the 0.75 cm value is, it appears to be a reasonable value from a clinical point of view, as repositioning the ETT is not a risk-free process.

We also tested separately the two mostly used anatomic markers “distance from the superior aspect of the right pulmonary artery” and “distance from the aortic arch apex. The first showed a sensitivity of 100% (95% CI: 97.4–100%), whereas specificity was not calculable following the absence of any true negative and false positive results. The latter showed a sensitivity of 89.35% (95% CI: 84.45–93.13%) and a specificity of 65.38% (95% CI: 44.33–82.79%).

Time Needed to Perform POCUS to Confirm ETT Position in the Newborns vs. the Reference Exams

Some of the studies selected in our research described the mean time needed to perform POCUS exams, with a variable range from seconds to minutes (Table 5 and Figure 6). A weighted average for the POCUS exams of about 10 min and 39 s can be described. Finally, some of these studies have also reported the time needed to perform the reference exams (CXR, Figure 6, or, in two cases, capnography). A weighted average to perform CXR can be estimated in 51 min.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Table 5. Time required for ETT position confirmation by POCUS and the reference exam.

What are the primary methods of confirming endotracheal tube placement within the trachea neonate?

Figure 6. Mean time (minutes) required for ETT confirmation by POCUS, compared with CXR (when performed and time information available). ETT, endotracheal tube; POCUS, point of care ultrasound; CXR, chest X-ray.

Discussion

This meta-analysis was novel in evaluating the accuracy and diagnostic value of POCUS for detecting ETT and assessing its placement. Previous meta-analysis studies described the diagnostic performance of POCUS regarding transient tachypnea of the newborn (TTN) (44, 45), respiratory distress syndrome (RDS) (46) and bronchopulmonary dysplasia (BPD) (47). The pooled results obtained from the selected studies demonstrated that POCUS can recognize ETT in 96.8% of cases and describe its correct placement with a pooled sensitivity of 93.44% ranging from 0.71 to 1.00.

Our comparison between POCUS and the standard technique (CXR) for the evidence of ETT and its correct placement in the neonatal population demonstrates that ultrasound examination has a good sensitivity. The findings of this meta-analysis confirm POCUS high diagnostic value concerning both items. Furthermore, the distance of ETT from the superior aspect of the RPA showed a higher sensitivity when compared with the distance from the AoA, whereas specificity was not calculable for the first marker. We believe that POCUS markers, intended as distance between ETT and RPA or between ETT and AoA, should be evaluated together in a multicentric study in order to determine the most accurate, fast and easy to perform the technique.

All the studies that considered radiography as the gold standard have shown that ultrasound is faster than radiography in confirming intubation and assessing the position of the ETT. However, some works analyzed the time needed for the ultrasound evaluation (since POCUS equipment was already available at the patient's crib) vs. the time needed for CXR to be performed, while others evaluated the activation time of the ultrasound service compared to the radiographic one from the moment of intubation. Therefore, when POCUS equipment was ready, ultrasound investigation proved to be very fast, from an average of 3 s [according to Takeuchi's work (41)] to <5 min in the other studies.

Regarding the assessment with capnography, this was faster compared to POCUS in one study (41) but slower in a second study (25). Possibly this difference can be justified by the population studied; indeed, the second study (25) considered only ELBW infants, a specific population where capnography can lead to false negative results or positivity latency in relation to low lung volumes (13).

It is important to consider several limitations with respect to the present meta-analysis. Patients' enrollment could not be consecutive in the great majority of the included studies since POCUS performers were not available 24/24 h and 7/7 days. Furthermore, personal experience with POCUS examination certainly plays a notable role in accuracy and speed of the echographic evaluation that we could not take into account in our analysis (40).

The 14 studies presented a low rate of ETT malposition (too high or too low inside the trachea) and an even lower rate of esophageal intubation, the latter probably evident to clinical evaluation, thus POCUS specificity in detecting ETT tip correct position is not reliable. Importantly, we must consider that a great number of ETT evaluations were not done immediately after the intubation procedure, but in already intubated and mechanically ventilated infants seeking to confirm the ETT (presumably adequately placed) and compare POCUS with CXR. In other cases (27, 39), POCUS was performed exclusively in well positioned ETT. We acknowledge that POCUS has some concerns regarding specificity but, considering the high sensibility and the high PPV, the impossibility of visualization of the ETT through this method could suggest a position that is too high. In this circumstance, probably, trying to deepen the ETT under ultrasound guidance could help.

Finally, we were not able to collect all data from all the studies: given the heterogeneity of the studies included in this meta-analysis, we tried to formulate homogeneous questions by requesting further data from the authors but we did not get all the answers.

Conclusions

One of the main challenges is to offer stronger evidence to promote the changing of clinical practice. Our meta-analysis shows that POCUS appears to be a fast and effective technique for identifying the appropriateness of ETT position in the NICU population.

More studies are needed in order to establish POCUS specificity, a large trial of POCUS performed immediately after intubation could offer a higher rate of malpositioned ETT and hence be more informative about specificity. POCUS training in ETT visualization in the newborn and comparison between different anatomical markers are topics worthy of further studies.

Author Contributions

SC, FS, MA, and LB: conceptualization, reviewing literature, drafting, writing, critically editing for important intellectual content, final approval of the version to be published, and agreeing to be accountable for all aspects of the work. SC, FS, and MA: reviewing literature, drafting, conducting statistical analysis, co-writing of the work. LB: solved questions related to the accuracy or integrity of any part of the work, reviewed the first draft of the paper. LB, SS, and DN: critically editing for important intellectual content, final approval of the version to be published. All authors contributed to the article and approved the submitted version.

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

We would like to thank the Departments of Pediatrics of Padua Hospital, the University of Padua and the Research strategy Program for pediatrics residents, for their support of research interests.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fped.2022.886450/full#supplementary-material

References

3. Alonso-Ojembarrena A, Lubián-López SP. Lung ultrasound score as early predictor of bronchopulmonary dysplasia in very low birth weight infants. Pediatr Pulmonol. (2019) 54:1404–9. doi: 10.1002/ppul.24410

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Singh Y, Tissot C, Fraga M V, Yousef N, Cortes RG, Lopez J, et al. International evidence-based guidelines on Point of Care Ultrasound (POCUS) for critically ill neonates and children issued by the POCUS Working Group of the European Society of Paediatric and Neonatal Intensive Care (ESPNIC). Crit Care. (2020) 24:1–16. doi: 10.1186/s13054-020-2787-9

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Sharma D, Tabatabaii SA, Farahbakhsh N. Role of ultrasound in confirmation of endotracheal tube in neonates: a review. J Matern Neonatal Med. (2019) 32:1359–67. doi: 10.1080/14767058.2017.1403581

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Hatch L, Grubb P, Lea A, Walsh W, Markham M, Whitney G, et al. Endotracheal intubation in neonates: a prospective study of adverse safety events in 162 infants. J Pediatr. (2016) 168:62–66.e6. doi: 10.1016/j.jpeds.2015.09.077

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Rivera R, Tibbals J. Complications of endotracheal intubation and mechanical ventilation in infants and children. Crit Care Med. (1992) 20:193–9. doi: 10.1097/00003246-199202000-00008

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Foglia EE, Ades A, Sawyer T, Glass KM, Singh N, Jung P, et al. Neonatal intubation practice and outcomes: an international registry study. Pediatrics. (2019) 143:1–10. doi: 10.1542/peds.2018-0902

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Rodríguez-Fanjul J, Balcells Esponera C, Moreno Hernando J, Sarquella-Brugada G. Lung ultrasound as a tool to guide the administration of surfactant in premature neonates. An Pediatría (English Ed). (2016) 84:249–53. doi: 10.1016/j.anpede.2015.09.034

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Schmölzer GM, O'Reilly M, Davis PG, Cheung PY, Roehr CC. Confirmation of correct tracheal tube placement in newborn infants. Resuscitation [Internet]. (2013) 84:731–7. doi: 10.1016/j.resuscitation.2012.11.028

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Madar J, Roehr CC, Ainsworth S, Ersdal H, Morley C, Rüdiger M, et al. European Resuscitation Council Guidelines 2021: Newborn resuscitation and support of transition of infants at birth. Resuscitation. (2021) 161:291–326. doi: 10.1016/j.resuscitation.2021.02.014

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Galiote JP, Ridoré M, Carman J, Zell L, Brant K, Gayle C, et al. Reduction in unintended extubations in a level IV neonatal intensive care unit. Pediatrics. (2019) 143:CD010221. doi: 10.1542/peds.2018-0897

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Fortas F, Loi B, Centorrino R, Regiroli G, Ben-Ammar R, Shankar-Aguilera S, et al. Enhanced INSURE (ENSURE): an updated and standardised reference for surfactant administration. Eur J Pediatr [Internet]. (2022) 181:1269–75. doi: 10.1007/s00431-021-04301-x

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, et al. Special report—neonatal resuscitation: 2010 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Pediatrics. (2010) 126. doi: 10.1542/peds.2010-2972E

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Sahu AK, Bhoi S, Aggarwal P, Mathew R, Nayer J, T A V., Mishra PR, Sinha TP. Endotracheal tube placement confirmation by ultrasonography: a systematic review and meta-analysis of more than 2500 patients. J Emerg Med [Internet]. (2020) 59:254–64. doi: 10.1016/j.jemermed.2020.04.040

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Chou EH, Dickman E, Tsou PY, Tessaro M, Tsai YM, Ma MHM, et al. Ultrasonography for confirmation of endotracheal tube placement: a systematic review and meta-analysis. Resuscitation [Internet]. (2015) 90:97–103. doi: 10.1016/j.resuscitation.2015.02.013

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Van de Voorde P, Turner NM, Djakow J, de Lucas N, Martinez-Mejias A, Biarent D, et al. European Resuscitation Council Guidelines 2021: paediatric life support. Resuscitation [Internet]. (2021) 161:327–87. doi: 10.1016/j.resuscitation.2021.02.015

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Descamps CS, Beissel A, Vo Van P, Butin M, Stacoffe N, Giai J, et al. Role of ultrasonography in the assessment of correct endotracheal tube placement in neonates. Acta Paediatr Int J Paediatr. (2020) 109:1057–9. doi: 10.1111/apa.15097

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Alonso Quintela P, Oulego Erroz I, Mora Matilla M, Rodríguez Blanco S, Mata Zubillaga D, Regueras Santos L. Usefulness of bedside ultrasound compared to capnography and X-ray for tracheal intubation. An Pediatría (English Ed). (2014) 81:283–8. doi: 10.1016/j.anpede.2014.01.002

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Zaytseva A, Kurepa D, Ahn S, Weinberger B. Determination of optimal endotracheal tube tip depth from the gum in neonates by X-ray and ultrasound. J Matern Neonatal Med. (2020) 33:2075–80. doi: 10.1080/14767058.2018.1538350

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Najib K, Pishva N, Amoozegar H, Pishdad P, Fallahzadeh E. Ultrasonographic confirmation of endotracheal tube position in neonates. Indian Pediatr. (2016) 53:886–8. doi: 10.1007/s13312-016-0953-6

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Reitsma JB, Leeflang MMG, Sterne JAC, Bossuyt PMM, Whiting PF, Rutjes AWSS, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. (2011) 155:529–36. doi: 10.7326/0003-4819-155-8-201110180-00009

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Sim S-S, Sun J-T, Fan C-M, Tsai K-C. The utility of ultrasonography to confirm proper endotracheal tube placement in neonates. Resuscitation. Irel. (2016) 106: e19–20. doi: 10.1016/j.resuscitation.2016.06.030

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Mora-Matilla M, Alonso-Quintela P, Oulego-Erroz I, Rodríguez-Blanco S, Gautreaux-Minaya S, Mata-Zubillaga D. Is ultrasound a feasible tool to verify endotracheal tube position in neonates? Resuscitation. Irel. (2013). 84:e19–20. doi: 10.1016/j.resuscitation.2012.09.026

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Oulego-Erroz I, Alonso-Quintela P, Rodríguez-Blanco S, Mata-Zubillaga D, Fernández-Miaja M. Verification of endotracheal tube placement using ultrasound during emergent intubation of a preterm infant. Resuscitation. Irel. (2012) 83:e143–4. doi: 10.1016/j.resuscitation.2012.02.014

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Galicinao J, Bush AJ, Godambe SA. Use of bedside ultrasonography for endotracheal tube placement in pediatric patients: a feasibility study. Pediatrics. (2007) 120:1297–303. doi: 10.1542/peds.2006-2959

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Chowdhry R, Dangman B, Pinheiro JMB. The concordance of ultrasound technique versus X-ray to confirm endotracheal tube position in neonates. J Perinatol. (2015) 35:481–4. doi: 10.1038/jp.2014.240

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Saul D, Ajayi S, Schutzman DL, Horrow MM. Sonography for complete evaluation of neonatal intensive care unit central support devices. J Ultrasound Med. (2016) 35:1465–73. doi: 10.7863/ultra.15.06104

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Salvadori S, Nardo D, Frigo AC, Oss M, Mercante I, Moschino L, et al. Ultrasound for endotracheal tube tip position in term and preterm infants. Neonatology. (2021) 118:569–77. doi: 10.1159/000518278

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Gorbunov A V., Koltunov I, Mazaev AP, Degtyareva M, Erokhina A, Demina A. The feasibility of ultrasound diagnostics in confirmation of endotracheal tube position in neonates. J Matern Neonatal Med. (2016) 29:1–313.

Google Scholar

39. Singh P, Thakur A, Garg P, Aggarwal N, Kler N. Normative data of optimally placed endotracheal tube by point-of-care ultrasound in neonates. Indian Pediatr. (2019) 56:374–80. doi: 10.1007/s13312-019-1533-3

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Sethi A, Nimbalkar A, Patel D, Kungwani A, Nimbalkar S. Point of care ultrasonography for position of tip of endotracheal tube in neonates. Indian Pediatr. (2014) 51:119–21. doi: 10.1007/s13312-014-0353-8

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Takeuchi S, Arai J, Nagafuji M, Hinata A, Kamakura T, Hoshino Y, et al. Ultrasonographic confirmation of endotracheal intubation in extremely low birthweigh infants—secondary publication. Pediatr Int. (2020) 62:175–9. doi: 10.1111/ped.14069

PubMed Abstract | CrossRef Full Text | Google Scholar

42. De Kock SH, Otto SF, Joubert G. The feasibility of determining the position of an endotracheal tube in neonates by using bedside ultrasonography compared with chest radiographs. South Afr J Child Heal [Internet]. (2015) 9:3–5. doi: 10.7196/sajch.740

CrossRef Full Text | Google Scholar

44. He L, Sun Y, Sheng W, Yao Q. Diagnostic performance of lung ultrasound for transient tachypnea of the newborn: a meta-analysis. PLoS ONE. (2021) 16:1–13. doi: 10.1371/journal.pone.0248827

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Wu J, Wang Y, Zhao A, Wang Z. Lung Ultrasound for the Diagnosis of Neonatal Respiratory Distress Syndrome: a Meta-analysis. Ultrasound Q. (2020) 36:102–10. doi: 10.1097/RUQ.0000000000000490

PubMed Abstract | CrossRef Full Text | Google Scholar

47. Pezza L, Alonso-Ojembarrena A, Elsayed Y, Yousef N, Vedovelli L, Raimondi F, De Luca D. Meta-analysis of lung ultrasound scores for early prediction of bronchopulmonary dysplasia. Ann Am Thorac Soc. (2022) 19:659–67. doi: 10.1513/AnnalsATS.202107-822OC

PubMed Abstract | CrossRef Full Text | Google Scholar